Setup Menu
Teledyne API – Model T300/T300M CO Analyzer
96
To set individual ranges press the following control button sequence.
SAMPLE
RANGE=50.0 PPM CO=XX.XX
<TST
TST> CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS
CLK MORE
EXIT
SETUP X.X
RANGE MODE MENU
MODE
SET UNIT DIL
EXIT
SETUP X.X
RANGE MODE:SNGL
SNGL DUAL
AUTO
DIL
ENTR EXIT
SETUP X.X
RANGE MODE:AUTO
SNGL DUAL AUTO
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT DIL
EXIT
SETUP X.X
HIGH RANGE:200.0 Conc
0
0
2
0
0
.0
ENTR EXIT
Toggle these buttons
to select the upper
SPAN
limit for the
reporting range.
EXIT
discards the new
setting.
ENTR
accepts the
new setting.
SETUP X.X
LOW RANGE:50.0 Conc
0
0
0
5
0
.0
ENTR EXIT
The
LOW
and
HIGH
ranges have separate
slopes and offsets for
computing the CO
concentration.
The two ranges must
be independently
calibrated
.
Avoid accidentally setting the
LOW
range (
RANGE1
) of the instrument
with a higher span limit than the
HIGH
range (
RANGE2
).
This will cause the unit to stay in
the low reporting range perpetually
and defeat the function of the
AUTO
range mode.
06864D DCN7562