Calibration Procedures
Teledyne API – Model T300/T300M CO Analyzer
200
9.3.2.
MANUAL CALIBRATION CHECKS WITH VALVE OPTIONS
INSTALLED
Wait until
STABIL
falls
below
0.2 PPM (for T300)
or 1.0 PPM (for T300M).
This may take several
minutes.
SAMPLE
STABIL= XXXX PPM
CO=XXXX
< TST TST > CAL
CALZ CALS
SETUP
The
ZERO
and/or
SPAN
buttons will appear at various
points of this process.
It is not necessary to press
them.
SAMPLE RANGE=50.0PPM CO=XX.XX
< TST
TST >
CAL CALZ
CALS
SETUP
Set the Display to show
the
STABIL
test function.
This function calculates
the stability of the CO
measurement
Toggle
TST>
button until ...
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until
STABIL
falls
below
0.2 PPM (for T300)
or 1.0 PPM (for T300M).
This may take several
minutes.
Allow span gas to enter the sample port
at the rear of the analyzer.
Front panel display
shows all of the
available gas
measurements
throughout this
procedure.
Record CO zero point readings
Record CO span point readings.