Calibration Procedures
Teledyne API – Model T300/T300M CO Analyzer
222
9.7.2.3.
ACTIVATE CO2 SENSOR STABILITY FUNCTION
To change the stability test function from CO concentration to the CO
2
sensor output,
press:
Press EXIT 3
times to return
to SAMPLE
menu.
SETUP X.X
STABIL_GAS:CO2
CO CO2 O2
ENTR
EXIT
SAMPLE RANGE=50.0 PPM CO= XX.XX
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG ALRM
EXIT
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
<PREV
NEXT>
JUMP
EDIT PRNT EXIT
SETUP X.X
2) STABIL_GAS=CO
<PREV NEXT> JUMP
EDIT
PRNT EXIT
SETUP X.X
STABIL_GAS:CO
CO
CO2
O2
ENTR EXIT
Continue pressing NEXT until ...
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
CO
2
and O
2
options only
appear if
associated
sensors are
installed.
Note
Use the same procedure to reset the STB test function to CO when the
CO
2
calibration procedure is complete.