![TELEDYNE API T300 Operation Manual Download Page 343](http://html.mh-extra.com/html/teledyne-api/t300/t300_operation-manual_1080880343.webp)
Teledyne API - T300/T300M and M300E/EM 04906J DCN7122
APPENDIX A-7: MODBUS Register Map
A-23
Table A-8: Terminal Key Assignments
TERMINAL KEY ASSIGNMENTS
ESC
Abort line
CR (ENTER)
Execute command
Ctrl-C
Switch to computer mode
COMPUTER MODE KEY ASSIGNMENTS
LF (line feed)
Execute command
Ctrl-T
Switch to terminal mode
APPENDIX A-7: MODBUS Register Map
Table A-9: MODBUS Register Map
MODBUS
Register Address
(dec., 0-based)
Description
Units
MODBUS Floating Point Input Registers
(32-bit IEEE 754 format; read in high-word, low-word order; read-only)
0
Detector measure reading
mV
2
Detector reference reading
mV
4
M/R ratio.
none
6
Slope for range #1
none
8
Slope for range #2
none
10
Offset for range #1
none
12
Offset for range #2
none
14
Concentration for range #1 during zero/span calibration, just before
computing new slope and offset
PPM
16
Concentration for range #2 during zero/span calibration, just before
computing new slope and offset
PPM
18
Concentration for range #1
PPM
20
Concentration for range #2
PPM
22
Concentration stability
PPM
24
Bench temperature
°
C
26
Bench temperature control duty cycle
Fraction
(0.0 = off,
1.0 = on full)
28
Wheel temperature
°
C
30
Wheel temperature control duty cycle
Fraction
(0.0 = off,
1.0 = on full)
32
Sample temperature
°
C
34
Sample pressure
“Hg
36
Internal box temperature
°
C
38
Photometer detector temperature drive
mV
06864D DCN7562