Adjustment Procedures
TLA Series Digitizing Oscilloscope Modules Service Manual
5- 11
This procedure adjusts the setup and hold parameters of the DSO module.
SW test names
SETUP_HOLD_CAL
Equipment
required
No change from previous test
Prerequisites
Warm-up time: 30 minutes
The logic analyzer application is not running
Test equipment connected as shown in Figure 5--2
Power-up diagnostics pass
Base Calibration and Pulse Trigger adjustment completed
1.
Verify that all of the prerequisites listed previously are met for the procedure.
2.
Follow the on-screen instructions to run the test for the DSO module.
3.
Verify no failures occur for each test. If a test continues to fail, check the
Troubleshooting
section on page 5--5.
DSO Procedure 4:
Setup and Hold
Adjustment
Summary of Contents for TLA Series
Page 4: ......
Page 12: ...Service Safety Summary viii TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 16: ...Preface xii TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 20: ...Introduction xvi TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 40: ...Theory of Operation 3 4 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 114: ...Repackaging Instructions 6 36 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 116: ...Options 7 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 118: ...Replaceable Electrical Parts 8 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 129: ......
Page 130: ......