Adjustment Procedures
5- 6
TLA Series Digitizing Oscilloscope Modules Service Manual
This procedure performs an internal and external calibration on the DSO module.
The
INTERNAL_CAL
test is the performance verification and adjustment software
equivalent of
SELF
_CAL.
SW test names
EXTERNAL_CAL, INTERNAL_CAL
Equipment
required
Voltage reference (item 7)
Precision BNC cable (item 12)
Dual banana-to-BNC adapter (item 6)
Capacitor, 0.1
F
1
(item 8)
Prerequisites
Warm-up time: 30 minutes
Test equipment connected as shown in Figure 5--1
Power-up diagnostics pass
No previous tests required
1
Install the 0.1
F capacitor across the voltage reference output terminals to reduce
noise.
DSO module
CH 1
--
+
Voltage reference
Dual
banana-to-BNC
adapter
Precision BNC
cable
0.1
F Capacitor
Figure 5- 1: Initial setup for the base calibration adjustment
DSO Procedure 1:
Base Calibration
Summary of Contents for TLA Series
Page 4: ......
Page 12: ...Service Safety Summary viii TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 16: ...Preface xii TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 20: ...Introduction xvi TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 40: ...Theory of Operation 3 4 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 114: ...Repackaging Instructions 6 36 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 116: ...Options 7 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 118: ...Replaceable Electrical Parts 8 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 129: ......
Page 130: ......