Performance Verification
4- 22
TLA Series Digitizing Oscilloscope Modules Service Manual
This procedure checks the DC accuracy for the internal and external trigger level
(DC coupled) of the DSO module.
SW test name
TRIG_MAIN_ACC
Equipment
required
Function generator (item 13)
Precision BNC cable (item 12)
Prerequisites
Warm-up time: 30 minutes
The logic analyzer application is not running
The performance verification and adjustment software is loaded
Test equipment connected as shown in Figure 4--8
Power-up diagnostics pass
INTERNAL_CAL test passes
Function generator
DSO module
CH 1
Precision BNC
cable
Figure 4- 8: Initial setup for the main trigger accuracy test
1.
Verify that all of the prerequisites listed previously are met for the procedure.
2.
Follow the on-screen instructions to run the test for the DSO module.
3.
Verify that all of the tests pass. If a test fails, check the
Troubleshooting
section on page 4--12.
DSO Procedure 9:
Main Trigger Accuracy
Summary of Contents for TLA Series
Page 4: ......
Page 12: ...Service Safety Summary viii TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 16: ...Preface xii TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 20: ...Introduction xvi TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 40: ...Theory of Operation 3 4 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 114: ...Repackaging Instructions 6 36 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 116: ...Options 7 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 118: ...Replaceable Electrical Parts 8 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 129: ......
Page 130: ......