Adjustment Procedures
TLA Series Digitizing Oscilloscope Modules Service Manual
5- 7
1.
If the logic analyzer application is running, quit the application.
2.
Load the performance verification and adjustment software as described in
the
Software Installation and Removal
on page 2--4.
3.
Verify that all of the prerequisites listed previously are met for the procedure.
4.
Run the performance verification and adjustment software as described in
Using the
Software
on page 4--9. Select TLA_dig, then select the correct
module type and the ADJ test option.
5.
Follow the on-screen instructions to run the
EXTERNAL_CAL
test for the DSO
module.
6.
Verify that
the EXTERNAL_CAL test
passes before disconnecting the test
equipment and performing the
INTERNAL_CAL test
.
7.
Verify that these tests pass before continuing with any other tests.
Summary of Contents for TLA Series
Page 4: ......
Page 12: ...Service Safety Summary viii TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 16: ...Preface xii TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 20: ...Introduction xvi TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 40: ...Theory of Operation 3 4 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 114: ...Repackaging Instructions 6 36 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 116: ...Options 7 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 118: ...Replaceable Electrical Parts 8 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 129: ......
Page 130: ......