Performance Verification
4- 10
TLA Series Digitizing Oscilloscope Modules Service Manual
The screen will display an error message if the DUT chosen does not match
the installed DUT.
7.
Click Enter to continue.
8.
Enter the complete serial number of the DUT (for example, B010100). Click
Enter to continue.
If you select no, a prompt asks you to enter the serial number again.
9.
The program lists sequences for PV (performance verification) and ADJ
(adjustments). Enter a number to select which sequence you want and click
Enter to continue.
10.
If an instrument is being tested, the program lists the different probe types
available for testing. Enter the appropriate number corresponding to your
probe and then click Enter to continue.
11.
Enter the operating temperature in degrees C (entries in the range of 20 to 30
degrees are valid). Click Enter to continue.
12.
Enter the operating humidity as a percentage (0% to 100% entries are valid).
Click Enter to continue.
13.
Determine which sequence to run:
H
RUN FULL SEQUENCE runs the entire sequence from beginning to
end. This is the recommended selection.
H
RUN PARTIAL SEQUENCE runs part of the full sequence. The
sequence runs from the selected starting point to the end of the sequence.
Summary of Contents for TLA Series
Page 4: ......
Page 12: ...Service Safety Summary viii TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 16: ...Preface xii TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 20: ...Introduction xvi TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 40: ...Theory of Operation 3 4 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 114: ...Repackaging Instructions 6 36 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 116: ...Options 7 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 118: ...Replaceable Electrical Parts 8 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 129: ......
Page 130: ......