Removal and Installation Procedures
TLA Series Digitizing Oscilloscope Modules Service Manual
6- 23
Use the following procedure and Figure 6--9 to install the Acquisition Board.
NOTE
. After replacing the Acquisition Board, you must verify operation and
perform the adjustment and Performance Verification procedures.
1.
Plug two (2) 100-pin ribbon cables onto J100 and J101. Be sure to firmly
seat the cables into their sockets so the cable retainers snap into place.
2.
Install the RF shielding gasket onto the BNC connectors.
3.
Perform the
Probe Interface
installation procedure (page 6--21).
Rear EMI Gaskets
You will want to use both hands to replace the four rear EMI Gaskets.
Use the following procedure to remove the rear EMI gaskets.
1.
Remove the cover (page 6--10).
2.
For the two rear gaskets on the chassis, remove the processor board
(page 6--13) and the acquisition board (page 6--22).
3.
Lift the gasket fingers and rotate the gasket off.
Lift gasket finger
1
Rotate
2
Figure 6- 10: Rear EMI gasket removal
Installation
Removal
Summary of Contents for TLA Series
Page 4: ......
Page 12: ...Service Safety Summary viii TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 16: ...Preface xii TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 20: ...Introduction xvi TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 40: ...Theory of Operation 3 4 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 114: ...Repackaging Instructions 6 36 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 116: ...Options 7 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 118: ...Replaceable Electrical Parts 8 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 129: ......
Page 130: ......