![Tektronix TLA Series Service Manual Download Page 22](http://html1.mh-extra.com/html/tektronix/tla-series/tla-series_service-manual_1078206022.webp)
Specifications
1- 2
TLA Series Digitizing Oscilloscope Modules Service Manual
Characteristics Tables
This section lists the specifications for the DSO module. All specifications are
guaranteed unless noted “typical.” Specifications that are marked with the
n
symbol are checked in the
Performance Verification
chapter of this manual.
The performance limits in this specification are valid with these conditions:
H
The DSO module must have successfully passed self calibration at an
ambient temperature b20
_
C and +30
_
C and after a warm-up period
of 30 or more minutes at these conditions.
H
The DSO module must be in an environment with temperature, altitude,
humidity, and vibration within the operating limits described in these
specifications. This includes the warm-up period of 30 or more minutes at
these conditions.
Table 1- 2: Signal acquisition system
Characteristic
Description
n
Accuracy, DC Gain
±
1.5% for full scale ranges from 20 mV to 100 V
±
2.0% for full scale ranges <19.9 mV
n
Accuracy, Internal Offset
1
Full Scale Range Setting
Offset Accuracy
10 mV -- 1 V
1.01 V -- 10 V
10.1 V -- 100 V
±
[(0.2%
×
| offset |) + 1.5 mV + (6%
×
full
scale range)]
±
[(0.25%
×
| offset |) + 15 mV + (6%
×
full
scale range)]
±
[(0.25%
×
| offset |) + 150 mV + (6%
×
full
scale range)]
n
Analog Bandwidth, DC--50
Ω
Coupled
Full Scale Range Setting
Bandwidth
2
10.1 V -- 100 V
DC -- 500 MHz (TLA7E1 and TLA7E2)
DC -- 500 MHz (TLA7D1 and TLA7D2)
100 mV -- 10 V
DC -- 1 GHz (TLA7E1 and TLA7E2)
DC -- 500 MHz (TLA7D1 and TLA7D2)
50 mV -- 99.5 mV
DC -- 750 MHz (TLA7E1 and TLA7E2)
DC -- 500 MHz (TLA7D1 and TLA7D2)
20 mV -- 49.8 mV
DC -- 600 MHz (TLA7E1 and TLA7E2)
DC -- 500 MHz (TLA7D1 and TLA7D2)
10 mV -- 19.9 mV
DC -- 500 MHz (TLA7E1 and TLA7E2)
DC -- 500 MHz (TLA7D1 and TLA7D2)
1
Net offset is the nominal voltage level at the DSO module input that corresponds to the center of the A/D Converter
dynamic range. Offset accuracy is the accuracy of this voltage level.
2
The limits given are for the ambient temperature range of 0
_
C to +30
_
C. Reduce the upper bandwidth frequencies by
5 MHz for each
_
C above +30
_
C. The bandwidth must be set to FULL.
Summary of Contents for TLA Series
Page 4: ......
Page 12: ...Service Safety Summary viii TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 16: ...Preface xii TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 20: ...Introduction xvi TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 40: ...Theory of Operation 3 4 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 114: ...Repackaging Instructions 6 36 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 116: ...Options 7 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 118: ...Replaceable Electrical Parts 8 2 TLA Series Digitizing Oscilloscope Modules Service Manual...
Page 129: ......
Page 130: ......