![Tektronix TAP2500 Instruction Manual Download Page 70](http://html1.mh-extra.com/html/tektronix/tap2500/tap2500_instruction-manual_1078181070.webp)
Performance veri
fi
cation
Test system rise time
1.
Connect the 80E04 sampling head to
Channel1 of the sampling oscilloscope.
2.
Connect the 80E0X sampling head to
Channel8 of the sampling oscilloscope.
3.
Connect SMA cables to Channels 1 and 8.
4.
Connect the SMA cable from Channel 1 to
the HF Probe Tip adapter.
52
TAP2500, TAP3500, and TAP4000 Instruction Manual