Speci
fi
cations
Warranted characteristics
Warranted characteristics describe guaranteed performance within tolerance limits or certain type-tested requirements. Warranted
characteristics that have checks in the
Performance Veri
fi
cation
section are marked with the
symbol. The Warranted, Typical,
and Mechanical characteristics apply to the TAP2500, TAP3500, and TAP4000 probes unless noted otherwise.
Table 1: Warranted electrical characteristics
Characteristic
Description
DC attenuation accuracy (probe only)
10:1 ±2% (excludes offset error)
Rise time (probe only)
≤
140 ps (TAP2500)
≤
130 ps (TAP3500)
Temperature
Operating: 0 to +50 _C (+32 to +122 °F), Nonoperating: -40 to +71 °C (-40 to
+160 °F)
Humidity
Operating: 5-95% RH, tested up to +30 °C (+86 °F) 5-85% RH, tested at +30 °C to
+50 °C (+86 to +122 °F)
Nonoperating: 5-95% RH, tested up to +30 °C (+86 °F) 5-85% RH, tested at
+30 °C to +75 °C (+86 to +167 °F)
Altitude
Operating: Up to 3000 meters (9,843 feet),
Nonoperating: Up to 12,000 meters (39,370 feet)
TAP2500, TAP3500, and TAP4000 Instruction Manual
37