Installation
AutoZero
We recommend that you run the probe
AutoZero routine:
After the 20 minute warm-up period
When the operating temperature of the
probe changes by ±5 °C
1.
Press the probe
Menu
button to display the
Probe Control screen on the oscilloscope.
2.
Short the probe tip to ground.
3.
Press the
AutoZero
button on the
instrument to execute the AutoZero routine.
TAP2500, TAP3500, and TAP4000 Instruction Manual