Performance veri
fi
cation
Equipment setup
Use the following procedure to set up and warm up the equipment to test the probe.
1.
Turn on the TekVPI oscilloscope.
2.
Connect the TekVPI Calibration/Veri
fi
cation
adapter to the oscilloscope.
3.
Connect the probe to the TekVPI
Calibration/Veri
fi
cation adapter and verify
that the Status LED on the probe turns
green.
4.
Turn on the remaining test equipment.
5.
Allow 20 minutes for the equipment to
warm up.
6.
Photocopy the test record and use it to
record the test results. (See page 59,
TAP2500, TAP3500, and TAP4000 Instruction Manual
47