Performance veri
fi
cation
8.
Set oscilloscope offset to 0.0 V.
9.
Set the DMM to DCV.
10.
Press the REL (relative) button to Zero the
DMM.
11.
Set the DC voltage source to +1.00 VDC
and enable the output.
12.
Record the DMM measurement in the test
record.
13.
Set the DC voltage source to -1.00 VDC.
14.
Record the DMM measurement in the test
record.
NOTE.
An unacceptable error value may
result if a precision 50
Ω
termination is not
used for the recommended termination, or if
the oscilloscope offset is not set to zero.
15.
Leave the TekVPI Calibration/Veri
fi
cation
adapter and probe connected the
oscilloscope to keep the probe warm.
50