Performance veri
fi
cation
DC gain accuracy
This test checks the DC gain accuracy of the probe.
1.
Connect the BNC-to-dual banana adapter
to the DMM input.
2.
Connect the SMA M-to-BNC F adapter
to the SMA output of the TekVPI
Calibration/Veri
fi
cation adapter.
3.
Connect the precision termination to the
BNC end of the SMA M-to-BNC F adapter.
4.
Connect the BNC cable between the
precision termination and the BNC-to-dual
banana adapter attached to the DMM.
48
TAP2500, TAP3500, and TAP4000 Instruction Manual