Calibration
Test procedure
Check the Calibration Status of the instrument:
1.
From the Utilities menu, select
Instrument
Calibration
.
2.
In the Calibration box, check that
Pass
is
displayed in the Status
fi
eld. If it is not,
disconnect all probes and signal sources
from the oscilloscope, and run the Signal
Path Compensation routine (see next step).
TAP2500, TAP3500, and TAP4000 Instruction Manual
9