Calibration Verification Test
C–6
ST112 SONET Transmission Test Set
bnc cable, a 50 ohms bnc thru terminator, and a bnc 50 ohms cable are needed. Leave everything
as it was in the last section. There are eleven steps in this section of the procedure.
1. On the front panel, press SETUP until the display shows the SONET Quickset menu.
2. Using the INC/DEC keys, select STSX1. Using the LEFT/RIGHT cursors, select DS3 and
then press ENTER.
3. Setting the signal around the center of the scope, measure the amplitdue of the signal.
Note:
The amplitude of the signal is its peak-to-peak value.
The amplitude of the STSX1 signal
should equal 1 Vpp + 200mV.
4. Compare the signal on the screen to the standard STS1 mask. Only the scope’s vertical
adjusment may be moved in order to make the signal fit within the mask.
Note: The
CSA803A oscilloscope contains the standard template in memory.
5. Set the vertical adjustment back to the normal position.
6. Connect the 50 ohms bnc cable to STS1 CLOCK OUT on the transmitter side of the rear
panel.
7. Connect the other end of the 50 ohms cable to the AC coupler.
8. Connect the AC coupler to one end of the 50 ohms thru terminator.
9. Connect the thru terminator to the frequency counter.
10. The STS1 frequency is 51.8400 Mhz + 51 Hz. The frequency shown on the counter should
be 51.8400 Mhz + 51 Hz.
11. Remove all cables and terminators from the unit and scope.
Note: For further information regarding the standard masks for DS1, DS3, and STS1 refer to
the ANSI technical standard, T1.102. The masks supported by the CSA803A scope are
direct representations of these standard masks.
OC1 Signal Mask Verification
This section of the procedure verifies that the output signal matches the standard OC1 mask. The
CSA803A scope, a single mode optical cable, an in-line 5db optical attenuator, an
Optical/Electrical converter, and a 50 ohms bnc cable are needed. For all optical sections, an in-
line 5db attenuator is required between the transmitter and receiver optical connectors. Before
any connections are made, use the optical cleaning kit to clean the cable end, the attenuator end,
and the optical connectors on the front panel of the unit. The last section of the procedure
provides an equipment list and simple instructions on how to clean the optical connectors. There
are seven steps in this section of the procedure.
1. Connect the bnc cable from the electrical end of the Optical/Electrical converter to the scope.
2. Connect one end of the optical cable to the optical connector on the transmitter side of the
front panel.
3. Connect the other end of the optical cable to the optical end of the Optical/Electrical
converter.