Incoming Inspection Test
ST112 SONET Transmission Test Set
B–17
Note: This procedure should be performed each time the fiber assemblies are to be used to
prevent scratching the fiber and ultimately destroying optical connectors.
1. Remove the protective optical cover from the connector to be cleaned.
2. Take a swab from the package and dip it into the clean Alcohol solution.
3. Gently insert the swab into the barrel (ferrule) opening of the fiber optic
connector until it stops.
4. Gently spin the swab clockwise and the counter clockwise.
5. Remove the swab.
6. Using the clean air spray, spray the air into the barrel opening for at least three
seconds.
Note:
Hold the can in an upright position.
DO NOT
tilt, invert,
or shake the can during use. If a mist appears from the can, the cleaning
process
must be performed again
.
7.
Do not
reuse the swab. Dispose of it immediately.
Cleaning Procedure for Unconnected Fiber Cables
Note: This procedure should be performed each time the fiber cables are to be used to prevent
scraching the fiber and ultimately destroying the cables.
1. Hold the fiber optic connector to be cleaned in your hand.
2. Try not to put unnecessary strain on the fiber.
3. Take a clean Kimwipe from the box and dab it into the clean Alcohol.
4. Wipe the Alcohol soaked Kimwipe over the ends of the fiber optic cable gently.
5. Using the clean air spray, spray the air into the end of fiber for at least three seconds.
Note:
Hold the can in an upright position.
DO NOT
tilt, invert, or shake the can during use. If a
mist appears from the can, the cleaning process
must be performed again
.
6. Dispose of the used Kimwipe immediately.
Do not
reuse it.