Incoming Inspection Test
ST112 SONET Transmission Test Set
B–7
15. Using the LEFT/RIGHT cursors, select D9.
16. Using the INC/DEC keys and the LEFT/RIGHT cursors, change D9 to 00 and then press
ENTER.
17. On the receiver side of the front panel, press OVHD and observe that the values of D7, D8,
and D9 are once again zero.
Continuous Alarms (STSX1)
This section of the procedure ensures that the continuous alarm settings are functioning properly.
Leave everything set-up as it was in the last section. There are thirty-five steps in this section of
the procedure.
1. On the transmitter side of the front panel, press ALARM until the display shows the Errored
Frame menu.
2. Using the LEFT/RIGHT cursors, select CONT and then press ENTER.
3. On the receiver’s STATUS section of the front panel, observe that the following error lights
appear. Note: If a red LED remains lit, then the error or alarm is still occuring. If a LED
under the POINTER heading remains lit, then the condition is still occuring.
A.
Under the SECTION heading, the red LEDs for LOF, OOF,
and B1 remain lit.
B.
Under the LINE heading, the red LEDs for AIS, LOP, RDI,
and B2 remain lit. The red LEDs for AIS, LOP, and RDI
may
flash.
C.
Under the PATH heading, the red LEDs for RDI, FEBE, and
B3 remain lit. The red LED for RDI may flash.
D.
Under the PAYLOAD/DS1/DS3 heading, the red LEDs for
LOF, PAR, SYNC, and BIT remain lit.
E.
Under the SPE section of the POINTER heading, the yellow
LED for NDF remains lit and the green LEDs for +PJ and
-PJ remain lit.
4. Using the LEFT/RIGHT cursors, select OFF and then press ENTER.
5. Press RUN/STOP twice to clear all errors.
6. On the transmitter side of the front panel, press ALARM until the display shows the Line-
RDI (FERF) menu.
7. Using the LEFT/RIGHT cursors, select CONT and then press ENTER.
8. On the receiver’s STATUS section of the front panel, under the LINE heading, observe that
the red LED for FERF remains lit.
9. Using the LEFT/RIGHT cursors, select OFF and then press ENTER.
10. Press RUN/STOP twice to clear all errors.