Incoming Inspection Test
B–16
ST112 SONET Transmission Test Set
11. On the display, the value of B1 equals three. However, because of the position of the error
being injected, B1 may equal up to six. For the same reason, the value of FEBE will be
either zero or one and the values of B2 and B3 will be between zero and six.
12. Press RUN/STOP twice to clear all errors.
13. On the transmitter side of the front panel, press ALARM until the display shows the Errored
Frame menu.
14. Using the LEFT/RIGHT cursors, select CONT and then press ENTER.
15. On the receiver’s STATUS section of the front panel, observe that the following error lights
appear.
A.
Under the SECTION heading, the red LEDs for LOF and OOF
remain
lit.
B.
Under the LINE heading, the red LEDs for AIS, LOP, and RDI remain lit.
The red LEDs for AIS, LOP, and RDI may flash.
C.
Under the PATH heading, the red LEDs for RDI, FEBE, and
B3 remain lit. The red LED for RDI may flash.
D.
Under the PAYLOAD/DS1/DS3 heading, the red LEDs for
LOF, PAR, SYNC, and BIT remain lit.
E.
Under the SPE section of the POINTER heading, the yellow
LED for NDF remains lit and the green LEDs for +PJ and
–PJ remain lit.
14. Using the LEFT/RIGHT cursors, select OFF and then press ENTER.
15. Press RUN/STOP twice to clear all errors.
You have now successfully verified that the unit is functional. If you had any problems with the
unit while performing the tests, repeat the procedure paying careful attention to the set-up of the
units and the directions given. If the unit still has a problem, please call Tektronix Customer
Sales and Support Center (CSSC) at 1-800-835-9433 x2400 or 503-627-2400 for assistance.
Cleaning Optical Cables
Required Equipment
Clean Air Spray - Dust Free (suggested: Chemtronics E - Series Ultrajet)
Isopropyl
Alcohol
70%
.035” Swab (suggested: Fisher Brand Cat.No. 14-959-77)
Kimwipes Laboratory Wipes (suggested: PN 34155)
Cleaning Procedure for Front Panel Optical Connectors
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com