Incoming Inspection Test
B–6
ST112 SONET Transmission Test Set
8. On the display, observe that SPE is at 500, +PJ has a one under it, -PJ has a one under it, and
NDF has a two under it.
9. Press RUN/STOP twice to clear all errors.
10. On the transmitter side of the front panel, press POINTER.
11. Using the LEFT/RIGHT cursors, select 500 under NDF.
12. Using the INC/DEC keys and the LEFT/RIGHT cursors, set the NDF to 522 and then press
ENTER.
13. Press RUN/STOP twice to clear all errors.
Transport Overhead/Path Overhead
This section of the procedure ensures that the editing function is working accurately. Leave
everything set-up as it was in the last section. There are seventeen steps in this section of the
procedure.
1. On the transmitter side of the front panel, press OVHD until the top line shows D7, D8, D9,
and Z3.
2. Using the LEFT/RIGHT cursors, select D7. Note: The brackets showing the selection do not
appear in the OVHD menu. The cursor flashes underneath the position of the current
selection.
3. Using the INC/DEC keys and the LEFT/RIGHT cursors, change D7 to AA and then press
ENTER.
4. Using the LEFT/RIGHT cursors, select D8.
5. Using the INC/DEC keys and the LEFT/RIGHT cursors, change D8 to 11 and then press
ENTER.
6. Using the LEFT/RIGHT cursors, select D9.
7. Using the INC/DEC keys and the LEFT/RIGHT cursors, change D9 to 05 and then press
ENTER.
8. On the receiver side of the front panel, press OVHD until the top line shows D7, D8, D9, and
Z3.
9. On the display, observe that D7, D8, and D9 all reflect the values you placed into the
transmitter (i.e. D7 = AA, D8 = 11, D9 = 05, and Z3 = 00).
10. On the transmitter side of the front panel, press OVHD.
11. Using the LEFT/RIGHT cursors, select D7.
12. Using the INC/DEC keys and the LEFT/RIGHT cursors, change D7 to 00 and then press
ENTER.
13. Using the LEFT/RIGHT cursors, select D8.
14. Using the INC/DEC keys and the LEFT/RIGHT cursors, change D8 to 00 and then press
ENTER.