Commands Listed in Alphabetical Order
Group
Source
Syntax
[SOURce:]VOLTage:LIMit:STATe {0|OFF|1|ON}
[SOURce:]VOLTage:LIMit:STATe?
Related Commands
[SOURce:]VOLTage:LIMit[:LEVel]
Arguments
0
or
OFF
turns off the maximum voltage limit.
1
or
ON
turns it on.
Examples
VOLTAGE:LIMIT:STATE ON
[SOURce:]VOLTage:TRIGgered[:IMMediate]
This command is used to set the voltage level for the trigger function. The
command queues the next setting for the currently selected channel. The units
are kV, V, mV or
μ
V. When the instrument receives its next trigger, the currently
selected channel will be set to the speci
fi
ed value. So, 8000000
μ
V will set the
voltage to 8 V on the front panel.
Group
Source
Syntax
[SOURce:]VOLTage:TRIGgered[:IMMediate] {<voltage>|MIN|MAX}
[SOURce:]VOLTage:TRIGgered[:IMMediate]?
Related Commands
Arguments
<voltage>
is a
fl
exible decimal number (NRf) that may be type NR1, NR2 or
NR3. It speci
fi
es a level between the minimum voltage and maximum nameplate
voltage level for the power supply.
MIN
sets the voltage to the minimum level (0 V).
MAX
sets the voltage to the maximum level.
Examples
VOLTAGE:TRIGGERED 4.5V
2-36
Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual