Status and Events
Messages and Codes
Error and event codes with negative values are SCPI standard codes. Error and
event codes with positive values are unique to the Series 2200 Programmable
Multichannel DC Power Supplies.
Table 3-9: No event messages
Code
Message
0
No events to report; queue empty
1
No events to report; new events pending *ESR?
Command Errors
The following table shows the command error messages generated by improper
syntax. Check that the command is properly formed and that it follows the rules
in the section on command Syntax.
Table 3-10: Command error messages (CME bit 5)
Code
Message
101
Design error: Too many numeric suf
fi
ces in Command Spec
110
No Input Command to parse
114
Numeric suf
fi
x is invalid value
116
Invalid value in numeric or channel list, e.g. out of range
117
Invalid number of dimensions in a channel list
120
Parameter of type Numeric Value over
fl
owed its storage
130
Wrong units for parameter
140
Wrong type of parameter(s)
150
Wrong number of parameters
160
Unmatched quotation mark in parameters (single/double)
165
Unmatched bracket
170
Command keywords were not recognized
180
No entry in list to retrieve
190
Too many dimensions in entry to be returned in parameters
191
Too many char
Execution Errors
The following table lists the execution errors that are detected during execution of
a command.
Table 3-11: Execution error messages (EXE bit 4)
Code
Message
–200
Execution error
–221
Settings con
fl
ict