Commands Listed in Alphabetical Order
Arguments
<Current>
is a
fl
exible decimal number (NRf) that may be type NR1, NR2 or
NR3. It speci
fi
es a level between the minimum current and maximum nameplate
current level for the power supply.
MIN
sets the current to the minimum level (0 A).
MAX
sets the current to the maximum level.
Returns
<NR2>
is the current setting in amperes.
Examples
CURR 3A
CURR 30mA
CURR MIN
CURR?
might return 2.0000, which would be the current setting in amperes.
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]
This command is used to set the current step value.
Group
Source
Syntax
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]
{<current level>}
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]?
Related Commands
[SOURce:]CURRent[:LEVel]:UP[:IMMediate][:AMPLitude]
]CURRent[:LEVel]:DOWN[:IMMediate][:AMPLitude]
Arguments
The current level in A, mA, or
μ
A.
[SOURce:]CURRent:TRIGgered[:IMMediate]
This command is used to set the current level for the trigger function. The
command queues the next setting for the currently selected channel. The units are
A, mA or uA. When the instrument receives its next trigger, the currently selected
channel will be set to the speci
fi
ed value.
Group
Source
Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual
2-27