Commands Listed in Alphabetical Order
Syntax
[SOURce:]CURRent:TRIGgered[:IMMediate] {<current>|MIN|MAX}
[SOURce:]CURRent:TRIGgered[:IMMediate]?
Related Commands
Arguments
<Current>
is a
fl
exible decimal number (NRf) that may be type NR1, NR2 or
NR3. It speci
fi
es a level between the minimum current and maximum nameplate
current level for the power supply.
MIN
sets the current to the minimum level (0 A).
MAX
sets the current to the maximum level.
Examples
CURRENT:TRIGGERED
1.1A
[SOURce:]CURRent[:LEVel]:UP[:IMMediate][:AMPLitude] (No Query Form)
This command is used to increase the current level
by a step. The stepping current can be set by the
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]
command.
Group
Source
Syntax
[SOURce:]CURRent[:LEVel]:UP[:IMMediate][:AMPLitude]
Related Commands
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]
[SOURce:]OUTPut:ENABle
This command enables or disables the current channel. This command performs
the same function as the “Channel Enable” selection in the menu.
Group
Source
Syntax
[SOURce:]OUTPut:ENABle
[SOURce:]OUTPut:ENABle?
2-28