Commands Listed in Alphabetical Order
DOWN
sets the voltage level to decrease a step.
DEF
is the default level (1 V).
Returns
The voltage in kV, V, mV, or uV.
[SOURce:]VOLTage:LIMit[:LEVel]
This command limits the maximum voltage that can be programmed on the power
supply. This command will apply the limit to the currently-selected channel and
corresponds to the front-panel Max Voltage setting that can be found under the
Protection Settings submenu.
Group
Source
Syntax
[SOURce:]VOLTage:LIMit[:LEVel] {<voltage>|MIN|MAX}
[SOURce:]VOLTage:LIMit[:LEVel]?
Related Commands
Arguments
<voltage>
is a
fl
exible decimal number that may be type NR1, NR2 or NR3.
It speci
fi
es the voltage limit setting, which can range from 0 to the maximum
nameplate voltage of the power supply.
MIN
sets the maximum voltage to the minimum level (0 V).
MAX
sets the maximum voltage to the maximum level (note that the maximum
level may be somewhat higher than the nameplate).
Examples
VOLTAGE:LIMIT:STATE 6V
[SOURCE:]VOLTAGE:LIMIT[:LEVEL]
? might return
30.1
, which is the
maximum voltage limit setting on the current channel.
[SOURce:]VOLTage:LIMit:STATe
This command turns the maximum voltage for the currently selected channel on or
off on the current channel. This limit corresponds to the Max Volt Set command
on the front panel of the instrument.
Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual
2-35