Giving Demos of MSO Features
MagniVu (MSO4000/MSO3000):
MagniVu provides timing resolution down to 60.6 ps (MSO4000)/121.2 ps (MSO3000) for
10,000 samples. The maximum MagniVu sample rate is 16.5 GS/s and the record length is 10,000 points centered on the
trigger. The main acquisition has a sample rate of 500 MS/s and a record length of up to 10M points. MagniVu is acquired
every acquisition. You can switch between the MagniVu record and the main record at any time, running or stopped.
Per channel threshold settings (MSO4000):
The MSO4000 allows the user to set a unique logic threshold for each channel.
This supports customer designs with multiple logic families, all on the same board.
Multiple buses:
The MSO4000 monitors up to four serial or parallel buses simultaneously. The MSO2000/MSO3000 monitors
two buses simultaneously.
Setup/Hold bus triggering:
The MSO models provide setup/hold time violation triggering across an entire parallel bus,
including any or all of the 16 digital channels and the 4 analog ones. You can trigger on all 20 analog and digital channels if
you use Aux In as the clock.
Parallel bus triggering:
The MSO models add user-de
fi
ned, logic triggering on a parallel bus. You can assign all of the 4
analog and 16 digital channels to de
fi
ne a logic pattern.
Deep memory:
The MSO4000 provides 10 M record length on each of the analog and digital channels, standard on all models.
The MSO3000 provides 5 M record length on each channel. The MSO2000 provides 1 M record length on each channel.
Waveforms/second:
The MSO4000/MSO3000 provides a 50,000 waveform/second waveform capture rate on the analog
channels. The MSO2000 provides a 5,000 waveform capture rate. A higher rate means less dead time and more probability
of detecting waveform anomalies.
80
Demo 2 Board Instruction Manual
Summary of Contents for DPO2000 Series