Locating Signals
Counter Output Bits
Board label.
CNT_OUT0: CNT_OUT6
Connector grid location.
B5, B6
Description.
These are the 7-bits of the binary counter. The LSB is CNT OUT0 at 625 KHz, that is, half of the counter input
clock. There are random setup and hold time changes of 500 ps on bits 2 (CNT OUT1) and 4 (CNT OUT3) of this bus. Use these
signals to demonstrate setup time and hold time bus triggering using MagniVu.
The Counter Output Bits and the Counter Clock signals are on eight adjacent sets of header pins for easy connection to a
digital probe.
350 ps Delayed Clock
Board label.
350_PS_DELAY_CLK
Connector grid location.
H6
Description.
This is a 156 kHz square wave signal and a delayed copy. Use the nominal 350 ps delay to demonstrate
the 60.6 ps timing resolution of the MSO4000.
Random Burst
Board label.
RNDM_BURST
Connector grid location.
C6