Maintenance
—7623/R7623
Service
Check
that the
signal is properly connected and that the
interconnecting
cables are not defective. Also, check the
power
source. The associated plug-in
units can be checked
for
proper operation
by substituting other units which are
known
to be operating
properly (preferably of the same
types). If the
trouble persists after substitution, the 7623 is
probably
at fault.
3.
Visual
Check.
Visually check the
portion of the
instrument
in which
the
trouble is
located. Many troubles
can
be located by visual indications such as unsoldered
connections,
broken
wires,
damaged
circuit boards,
damaged components, etc.
4.
Check
Instrument
Calibration.
Check
the calibration
of
this
instrument, or the affected circuit if
the trouble
appears
in one circuit. The apparent trouble may only be a
result
of misadjustment, or
may be corrected by calibra
tion. Complete calibration
instructions are
given in the
Calibration
section.
5. Isolate
Trouble To
a Circuit. To isolate trouble to a
particular
circuit, note the trouble symptom. The symptom
often
identifies the circuit in which the trouble is
located.
For
example, poor focus indicates that the CRT
circuit
(includes high-voltage
supplies) is probably at
fault. When
trouble
symptoms appear in more than one circuit, check
affected
circuits by taking voltage
and waveform readings.
Typical
voltages and waveforms are given on the schematics
in the Diagrams section.
TABLE
4-2
Power
Supply Tolerance and
Ripple
Power
Supply
Test
Point
Output
Voltage
Tolerance
Maximum
ripple
(peak-to-peak)
-50
Volt TP -50 (back of
Main
Interface
board) on
P1170-Pin
1
±0.1
volt
5
mV
-15 Volt
P1170-Pin8
±0.3
volt
2
mV
+5 Volt
P1170-Pin 6
±0.15
volt
2
mV
+
15 Volt
P1170-Pin 5
±0.3
volt
2
mV
+50 Volt
P1170-Pin4
±0.6
volt
5
mV
+
130 Volt
P1170-Pin 3
±5.2 volts
300
mV
If
incorrect operation of the power supplies is suspected,
connect the
7623 to a variable autotransformer. Then,
check
for
correct regulation with a
DC voltmeter (0.1%
accuracy)
and correct ripple with
a test oscilloscope while
varying
the autotransformer throughout the regulating
range of this instrument.
After the
defective circuit
has been located, proceed
with steps 6 and 7 to locate the defective component(s).
6.
Check
Voltages and
Waveforms. Often the defective
component
can be located by checking for the correct
voltage
or waveform in the circuit. Typical voltages and
waveforms are given on the diagrams.
Incorrect
operation
of
all circuits often
indicates trouble
in the power supply. Check first for correct
voltage of the
individual
supplies. A defective component elsewhere in the
instrument
can
appear
as a power-supply trouble and may
also
affect the operation of other circuits. Table 4-2 lists
the
tolerances of the power supplies in this instrument.
These
voltages
are measured between the power-supply test
points
(see
Section
2 for test-point location) and ground. If
a
power-supply voltage is within
the listed tolerance, the
supply
can
be assumed to be working correctly. If outside
the
tolerance, the supply may be misadjusted or operating
incorrectly.
Use
the procedure given in the Calibration
section to adjust the power supplies.
Fig.
4-6 provides a guide for
locating a defective circuit.
This
chart
does not include checks for
all possible defects;
use
steps 6 and 7 in such
cases. Start from the top of the
chart
and perform the given
checks on the left side of the
page
until a step is found which does not produce the
indicated
results. Further checks and/or
the circuit in which
the trouble
is probably located are listed to the right of this
step.
NOTE
Voltages
and waveforms given on the diagrams are
not
absolute and may vary slightly between instru
ments.
To obtain operating conditions similar to
those
used to take
these readings, see the first
diagram
page.
7.
Check
Individual
Components.
The following
procedures
describe
methods of checking individual
components
in the 7623. Components which
are soldered
in place
are best checked by first disconnecting one end.
This
isolates
the measurement from the effects of
surrounding
circuitry.
A.
SEMICONDUCTORS.
Power switch must be turned off
before removing or
replacing
semiconductors.
4-8
Summary of Contents for 7623
Page 1: ...MANUAL 7623 R7623 STORAGE OSCILLOSCOPE SERVICE MANUFACTURERS OF CATHODE RAY OSCILLOSCOPES ...
Page 51: ...Fig 3 2 Block diagram of Logic circuit Circuit Description 7623 R 7623 Service ...
Page 72: ...W NJ 00 Fifl 3 22 Low Voltage Power Supply detailed block diagram ...
Page 73: ...Circuit Description 7623 R 7623 Service ...
Page 74: ...CO NJ CD Fig 3 22 Low Voltage Power Supply detailed block diagram cont ...
Page 75: ...Circuit Description 7623 R 7623 Service ...
Page 97: ...Circuit Description 7623 R7623 Service 3 51 ...
Page 98: ...Circuit Description 7623 R7623 Service Fig 3 39 Output Pulses for the Storage Circuits 3 52 ...
Page 103: ...Circuit Description 7623 R7623 Service 3 57 ...
Page 108: ... Ç À Fig 4 2 Location of circuit boards in the 7623 ...
Page 109: ...Fig 4 3 Location of circuit boards in the R7623 Maintenance 7623 R 7623 Service ...
Page 113: ...Maintenance 7623 R7623 Service Fig 4 6 Circuit Isolation Troubleshooting Chart 4 9 ...
Page 165: ...7623 BLOCK DIAGRAM ...
Page 166: ...7623 R7623 Service Front of Board ...
Page 167: ......
Page 168: ...FL ...
Page 169: ......
Page 173: ...7623 Logic ...
Page 175: ...Vertical Interface A4 ...
Page 178: ...Vertical Interface ...
Page 180: ...Vertical Amp A5 ...
Page 184: ...Horizontal Amp A6 ...
Page 186: ...7623 TO P450 VERT AMP 3 HORIZONTAL AMPLIFIER ...
Page 188: ...Output Signals A7 ...
Page 190: ...FROM 7G23 Output Signals g ...
Page 195: ...FROM LV POWER SUPPLY 7623 CRT CIRCUIT ...
Page 197: ......
Page 200: ......
Page 202: ...Storage Output A14 ...
Page 204: ......
Page 205: ...7623 R7623 Service Fig 6 14 A15 Cal Storage circuit board ...
Page 208: ......
Page 209: ...7623 R7623 Service Fig 6 15 A16 Readout System circuit board ...
Page 213: ......
Page 242: ...7623 R7623 OSCILLOSCOPE b ...
Page 243: ...151 ...
Page 246: ...20 ...
Page 247: ...7623 R7623 OSCILLOSCOPE 112 ...