VOLTAGES
AND
WAVEFORMS
The
voltages and
waveforms shown on this diagram were obtained using the recommended test equipment and test set-ups
listed
below.
RECOMMENDED
TEST
EQUIPMENT
ITEM
SPECIFICATIONS
RECOMMENDED
TYPE
Oscilloscope
Frequency
response
Deflection
factor
Input
impedance
Sweep
rate
DC
to
65 MHz
5
mV to 5 V/Div
10
MΩ,20
pF
500
ns
Tektronix
7603 or 7613
equipped
with
7A22
Amplifier
and 7B50
Time-Base
unit,
or equivalent.
Probe
Fast
rise
10X attenuation probe com
patible
with
the vertical amplifier of the
test
oscilloscope.
Tektronix
P6053A, or
equivalent.
Voltmeter (Non-Loading
Digital
Multimeter)
Input
impedance
Range
10 MΩ
0
-
500 V
Tektronix
7D13 Digital Multimeter (test
oscilloscope
must have readout system)
or
Fairchild Model 7050, or equivalent.
Calibration
Fixture
Used
for loading power
supply for ripple
waveforms.
Tektronix
Signal
Standardizer, Tektronix
Part
Number
067-0587-01, or equivalent.
Voltage
Measurements
Voltage
measurements
on this
diagram were made under
the following conditions:
Set
knob
type front
panel controls (except READ
OUT)
to mid-range.
Turn
READOUT to off.
Set
VERT MODE for CHOP.
Set for NON STORE
condition.
Set TRIG SOURCE for VERT
MODE.
No plug-ins
are installed.
Waveforms
Waveforms
shown on this diagram
were obtained under
the following conditions:
7623
OSCILLOSCOPE
UNDER TEST.
Front
panel
controls are
set the same as for voltage measurements. A
signal
standardizer
plug-in
unit is installed
in the vertical
plug-in
position
and
a
7B53AN Time Base unit in a
horizontal
position. The signal standardizer
plug-in unit
is used for
power supply loading
and is set for AUX
IN
with
a REP
RATE of 100 Hz.
No signal is applied. The
7B53AN
is set for
free running sweep of 1 ms/Division.
TEST OSCILLOSCOPE. The test oscilloscope
is AC
coupled
and set
for LINE trigger. A 7A22 Vertical
plug-in
unit is installed and set as follows:
HF-3 dB POINT
30
kHz
LF
-3 dB POINT
DC
+INPUT
AC
-INPUT
GND
Voltmeter common
is connected to chassis ground.
Tolerances of voltages and
waveforms shown are
20%.
Summary of Contents for 7623
Page 1: ...MANUAL 7623 R7623 STORAGE OSCILLOSCOPE SERVICE MANUFACTURERS OF CATHODE RAY OSCILLOSCOPES ...
Page 51: ...Fig 3 2 Block diagram of Logic circuit Circuit Description 7623 R 7623 Service ...
Page 72: ...W NJ 00 Fifl 3 22 Low Voltage Power Supply detailed block diagram ...
Page 73: ...Circuit Description 7623 R 7623 Service ...
Page 74: ...CO NJ CD Fig 3 22 Low Voltage Power Supply detailed block diagram cont ...
Page 75: ...Circuit Description 7623 R 7623 Service ...
Page 97: ...Circuit Description 7623 R7623 Service 3 51 ...
Page 98: ...Circuit Description 7623 R7623 Service Fig 3 39 Output Pulses for the Storage Circuits 3 52 ...
Page 103: ...Circuit Description 7623 R7623 Service 3 57 ...
Page 108: ... Ç À Fig 4 2 Location of circuit boards in the 7623 ...
Page 109: ...Fig 4 3 Location of circuit boards in the R7623 Maintenance 7623 R 7623 Service ...
Page 113: ...Maintenance 7623 R7623 Service Fig 4 6 Circuit Isolation Troubleshooting Chart 4 9 ...
Page 165: ...7623 BLOCK DIAGRAM ...
Page 166: ...7623 R7623 Service Front of Board ...
Page 167: ......
Page 168: ...FL ...
Page 169: ......
Page 173: ...7623 Logic ...
Page 175: ...Vertical Interface A4 ...
Page 178: ...Vertical Interface ...
Page 180: ...Vertical Amp A5 ...
Page 184: ...Horizontal Amp A6 ...
Page 186: ...7623 TO P450 VERT AMP 3 HORIZONTAL AMPLIFIER ...
Page 188: ...Output Signals A7 ...
Page 190: ...FROM 7G23 Output Signals g ...
Page 195: ...FROM LV POWER SUPPLY 7623 CRT CIRCUIT ...
Page 197: ......
Page 200: ......
Page 202: ...Storage Output A14 ...
Page 204: ......
Page 205: ...7623 R7623 Service Fig 6 14 A15 Cal Storage circuit board ...
Page 208: ......
Page 209: ...7623 R7623 Service Fig 6 15 A16 Readout System circuit board ...
Page 213: ......
Page 242: ...7623 R7623 OSCILLOSCOPE b ...
Page 243: ...151 ...
Page 246: ...20 ...
Page 247: ...7623 R7623 OSCILLOSCOPE 112 ...