Appendix B: Performance Verification
DG2040 Data Generator
BĆ43
3. Use the sampling oscilloscope measurement functions to perform the clock
signal rise and fall time measurement under the clock frequency settings
shown in Table B-8.
4. Change the SMA cable connection from the CLOCK and CLOCK outputs to
CH0 and CH0 outputs, respectively, and repeat step 3 for each of these
channels. Verify that the rise and fall times are within the specified ranges.
5. Change the SMA cable connection from the CH0 and CH0 outputs to CH1
and CH1 outputs, respectively, and repeat step 3 for each of these channels.
Verify that the rise and fall times are within the specified ranges.
Table BĆ8: Rise and fall time accuracies
Voltage settings
Low level
High level
Clock frequency
Rise and fall time limit
0 V
+1.00 V
100 MHz
t
150 psĂ(20Ă% -Ă80Ă%)
0 V
+1.00 V
500 MHz
t
150 psĂ(20Ă% -Ă80Ă%)
0 V
+1.00 V
1100 MHz
t
150 psĂ(20Ă% -Ă80Ă%)
6. Press the START/STOP button on the front panel to stop output (the
button’s LED goes off).
This completes the performance verification procedures.
Test Record
Photocopy the following 4 pages and use them to record the performance test
results for your DG2040 Data Generator.
Summary of Contents for DG2040
Page 1: ...User Manual DG2040 Data Generator 071 0257 04 This document supports firmware version 1 00...
Page 4: ......
Page 18: ...Preface xiv DG2040 Data Generator...
Page 19: ...Getting Started...
Page 20: ......
Page 32: ...Getting Started 1 12 DG2040 Data Generator...
Page 33: ...Operating Basics...
Page 34: ......
Page 87: ...Reference...
Page 88: ......
Page 152: ...Reference 3 64 DG2040 Data Generator...
Page 186: ...Reference 3 98 DG2040 Data Generator...
Page 187: ...Appendices...
Page 188: ......
Page 200: ...Appendix A Specifications A 12 DG2040 Data Generator...
Page 248: ...Appendix B Performance Verification B 48 DG2040 Data Generator...
Page 258: ...Appendix C Miscellaneous C 10 DG2040 Data Generator...
Page 259: ...Index...
Page 260: ......
Page 265: ......
Page 266: ......