Appendix B: Performance Verification
BĆ14
DG2040 Data Generator
This test verifies that the external reference input operation is functioning
correctly.
The following equipment is required.
Equipment Required
Oscilloscope
Function generator
Four BNC cables
Two SMA cables
Two SMA femaleĆtoĆBNC male adapters
Performance check disk
Connections.
Refer to Figure B-6 for connections.
Use the SMA cables and the SMA female-to-BNC male adapters to make the
following connections:
H
Connect CH1 output from the DG2040 Data Generator front panel to the
oscilloscope CH1 input.
H
Connect CH1 output to the DG2040 Data Generator front panel to the
oscilloscope CH2 input.
Use the BNC cables to make the following connections:
H
Connect the function generator CH1 output to the DG2040 Data Generator
rear panel event input.
H
Connect the function generator CH2 output to the DG2040 Data Generator
rear panel reference input.
H
Connect the event output from the DG2040 Data Generator rear panel to the
oscilloscope CH3 input.
H
Connect the sync output from the DG2040 Data Generator rear panel to the
oscilloscope CH4 input.
External Reference Input
Summary of Contents for DG2040
Page 1: ...User Manual DG2040 Data Generator 071 0257 04 This document supports firmware version 1 00...
Page 4: ......
Page 18: ...Preface xiv DG2040 Data Generator...
Page 19: ...Getting Started...
Page 20: ......
Page 32: ...Getting Started 1 12 DG2040 Data Generator...
Page 33: ...Operating Basics...
Page 34: ......
Page 87: ...Reference...
Page 88: ......
Page 152: ...Reference 3 64 DG2040 Data Generator...
Page 186: ...Reference 3 98 DG2040 Data Generator...
Page 187: ...Appendices...
Page 188: ......
Page 200: ...Appendix A Specifications A 12 DG2040 Data Generator...
Page 248: ...Appendix B Performance Verification B 48 DG2040 Data Generator...
Page 258: ...Appendix C Miscellaneous C 10 DG2040 Data Generator...
Page 259: ...Index...
Page 260: ......
Page 265: ......
Page 266: ......