Issue/Rev. 1.4 (9/18) ║ MNF18001EN / DOK-479-E • Page 85
3132136 – Correction
A fixed correction factor (= K-factor) can be set here, if a linear deviation
between the display on the MultiLevel and bell prover is determined during
measurements in the bell prover.
K affects only the volume in accordance with the level sensor table, not the
correction volume of the inclination table and not the residual volume!
V
target
= volume in the bell prover
V
actual
= MultiLevel display
K
old
= K-factor used to determine V
actual
.
V
target
x K
old
K
new
= ----------------
V
actual
ATTENTION:
Observe compensation, i.e. compare only V
T
or only V
0
!
3132137 – Preliminary Switch-Off
In the case of unfavourably shaped measuring compartments with large
residual volumes that can no longer be measured by the level measuring
instrument, there is a possibility to activate a preliminary switch-off. This
serves as an instruction to the driver to check whether there is enough
room in the storage tank for the entire residual volume. If this is not the
case, a printout of the partial volume already measured can be generated
and the remaining product metered out into another tank.
→ value in
mm
→ The value must be within a range where the level sensor can still measure
before the residual volume discharge. (With a standard setting, the value should
be > 45 mm.)
→ value =
0
means that the preliminary switch-off is deactivated.
3132138 – Inclination Stop
If the inclination of the vehicle is not within the permitted calibration limits when
the wet leg sensor runs dry, an uncalibrated measurement takes place, since it
is not certain whether residual quantities have remained in the compartment.
A preliminary check can be activated via the inclination stop. If the inclination
is not adhered to when the level falls below the set height, the delivery
is STOPPED and a corresponding error message is generated. The operator
can then correct the vehicle alignment if necessary and finish the delivery as
a calibrated delivery.
→ value in
mm
→ value =
0
means that the inclination stop is deactivated
→ value =
9999
means check before starting delivery
MultiLevel Instruction Manual
MENU structure
Summary of Contents for Sening MultiLevel
Page 6: ...Page 6 MNF18001EN DOK 479 E Issue Rev 1 4 9 18 Page intentionally left blank ...
Page 22: ...Page 22 MNF18001EN DOK 479 E Issue Rev 1 4 9 18 Page intentionally left blank ...
Page 36: ...Page 36 MNF18001EN DOK 479 E Issue Rev 1 4 9 18 Page intentionally left blank ...
Page 52: ...Page 52 MNF18001EN DOK 479 E Issue Rev 1 4 9 18 Page intentionally left blank ...
Page 131: ...Issue Rev 1 4 9 18 MNF18001EN DOK 479 E Page 131 MultiLevel Instruction Manual Form Layout ...
Page 138: ...Page 138 MNF18001EN DOK 479 E Issue Rev 1 4 9 18 Page intentionally left blank ...
Page 168: ...Page 168 MNF18001EN DOK 479 E Issue Rev 1 4 9 18 Page intentionally left blank ...