![tecan infinite 200 Instructions For Use Manual Download Page 80](http://html1.mh-extra.com/html/tecan/infinite-200/infinite-200_instructions-for-use-manual_828982080.webp)
5. Optical System
80
Instructions for Use for
infinite® 200
No. 30017581 Rev. No. 1.4
2008-07
5.5.3
Cuvette types
The cuvette port is compatible with the following cuvettes:
Cuvette Type
Width x Depth
Maximum Height
(including lid)
Filling Volume
Example
Standard cuvettes
12.5 x 12.5 mm
55 mm
2 ml
Hellma 110 QS,
10 mm
Semi-macro
cuvettes
12.5 x 12.5 mm
55 mm
1 ml
Hellma 108-QS,
10 mm
∗
Micro cuvettes
12.5 x 12.5 mm
55 mm
0.5 ml
Hellma 104.002
QS, 10 mm
∗
Ultra-micro
cuvettes
12.5 x 12.5 mm
55 mm
100 µl
Hellma 105.202,
10 mm
∗
Cuvettes with a measurement window < 2 mm (diameter) cannot be used.
STOP
Caution
Always use a valid filling volume. Make sure that the liquid level in the
cuvette exceeds 20 mm (height). Otherwise the light path in the cuvette
might not be filled completely with liquid which can lead to wrong
measurement results.
STOP
Caution
The cuvette port of the
infinite
®
M200
cannot be used for
cuvettes with a measurement window < 2 mm (diameter)
and a center height below 15 mm.
∗
Hellma GmbH & Co. KG, Germany; www.hellma-worldwide.com