2. General Description
44
Instructions for Use for
infinite® 200
No. 30017581 Rev. No. 1.4
2008-07
Example 2: Enliten
®
ATP Assay System Bioluminescence
Detection Kit for ATP (Promega Corp.)
For assay details please refer to www.promega.com.
Plate
Select an appropriate plate type. For luminescence
measurements, white microplates are recommended. For
this example, a white 96 well plate was selected.
Part of the plate
(Not shown); can be optionally selected if only part of the
plate should be processed
Well
Mandatory for measurements with ‘injection’
Injection
Injector A injects 100 µl with speed 100 µl/sec., refill mode:
standard
Wait (Timer)
2 s wait time
Luminescence
Luminescence measurement
with 10 s integration time, attenuation ‘none’
Move Plate
Plate is moved out after finishing all wells