Picture 2-1: Inserting the injector carrier into the injector port
STOP
Caution
The injector carrier must be in the service position
for washing und priming.
Prime and Wash must not be performed
when the injector is in the instrument!
STOP
Caution
If the injector carrier is not inserted correctly in the injector port, the
injector sensor will not detect the inserted injector and therefore
washing and priming will be enabled, which can damage the instrument.