8. Quality Control
138
Instructions for Use for
infinite® 200
No. 30017581 Rev. No. 1.4
2008-07
8.3.4
Fluorescence Polarization (
infinite
®
F200
only)
For the
infinite
®
F200
with the option ‘Fluorescence Polarization’ the following tests
may be performed to prove the specifications:
•
Precision/Repeatability
These test instructions are valid for the
infinite
®
F200
:
•
Standard version
•
Spectrally enhanced version
Precision
Perform the following measurement to determine the precision:
Measurement Parameters:
Parameter Setting
Reading Mode
Fluorescence Polarization
Exc Wavelength
485 (20 nm)
Em Wavelength
535 (25 nm)
Number of flashes
25
Integration Time
40
Lag Time
0
Time between Move and Read
0
Gain Optimal
Plate Type
GRE96fb
Reference from/to
A1 – D1
Reference blank from/to
A2 – D2
Reference value
20 mP
Measurement blank from/to
same as reference blank