Key Pump Featur
es
2
Chapter 7 – Bolus
4.
Tap
NEXT
to confirm the units of
insulin to be delivered.
You can always tap
View Cal-
culation
to display the Delivery
Calculation screen.
5.
Confirm Request.
Tap
YES
if entered data is correct.
Tap
NO
to go back to make
changes or view calculations.
6.
Tap
DELIVER
.
7.
The bolus initiated screen is
temporarily displayed.
7.5
Extended Bolus
The Extended Bolus feature allows you
to deliver part of the bolus now and
part of the bolus slowly over a period
of up to 8 hours. This can be helpful for
high fat meals such as pizza or if you
have gastroparesis (delayed stomach
emptying). When using extended bolus,
any correction bolus amount will always
be given in the DELIVER NOW portion.
Talk with our healthcare provider to
determine if this feature is appropriate
for you as well as for recommendations
on the split between now and later and
the duration for the later portion.
Only 1 extended bolus can be active
at any given time. However, if the
DELIVER LATER portion of an extend-
ed bolus is active, you can request
another standard bolus.
1.
From the Home Screen,
tap
BOLUS
.
2.
Tap
0 grams
(or 0
units
).
3.
Using the onscreen keypad enter
grams of carb (or units of insulin).
Tap
DONE
.
4.
If desired, tap
Add BG
and using
the onscreen keypad enter BG
value. Tap
DONE
.
5.
Tap
NEXT
to confirm the units of
insulin to be delivered.
You can always tap
View Cal-
culation
to display the Delivery
Calculation screen.
6.
Confirm Request.
Tap
YES
if entered data is correct.
Tap
NO
to go back to make
changes or view calculations.
7.
Tap
EXTENDED
to turn on the
extended feature, then tap
NEXT
.
8.
Tap
50%
under DELIVER NOW
to adjust the percentage of the
food bolus that is to be delivered
immediately.
The percentage value for DELIVER
LATER is automatically calculated
by the pump. The default is 50%
NOW and 50% LATER. The de-
fault for DURATION is 2 hours.
Summary of Contents for t:slim X2
Page 2: ......
Page 14: ...Section 1 Before You Begin...
Page 15: ...Chapter 1 Introduction...
Page 19: ...Chapter 2 Important Safety Information...
Page 31: ...Chapter 3 Getting to Know Your t slim X2 Pump...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profiles...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim X2 Pump Settings...
Page 109: ...Chapter 12 t slim X2 Pump Info and History...
Page 111: ...Chapter 13 t slim X2 Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim X2 Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim X2 Pump Alerts...
Page 143: ...Chapter 16 t slim X2 Pump Alarms...
Page 157: ...Chapter 17 t slim X2 Pump Malfunction...
Page 160: ...Section 5 Living With and Caring for Your t slim X2 Pump...
Page 161: ...Chapter 18 Lifestyle Issues and Travel...
Page 165: ...Chapter 19 Taking Care of Your t slim X2 Pump...
Page 168: ...Section 6 Technical Specifications and Warranty...