Customer Technical Support:
(877) 801-6901
64
Chapter 5 – Infusion Site Care and Loading Cartridge
5.2
Cartridge Instructions
for Use
For complete cartridge labeling, consult
the Cartridge Instructions for Use
included in the t:slim Cartridge box.
5.3
Filling and Loading a
t:slim Cartridge
This section describes how to fill the
cartridge with insulin and load the
cartridge into your t:slim X2 Pump. The
single-use disposable cartridge can
hold up to 300 units (3.0 mL) of insulin.
WARNING
DO NOT
use any other insulin with your pump
other than U-100 Humalog
®
or NovoLog
®
. Only
Humalog
®
and NovoLog
®
have been tested
and found to be compatible for use with this
Pump. Use of insulin with lesser or greater
concentration can result in under delivery or
over delivery of insulin. This can cause very
high or a very low blood glucose.
WARNING
DO NOT
reuse cartridges or use cartridges
other than those manufactured by Tandem
Diabetes Care, Inc. Use of cartridges not
manufactured by Tandem Diabetes Care, Inc. or
reuse of cartridges may result in over delivery
or under delivery of insulin. This can cause very
low or very high blood glucose.
Before you begin, make sure you
have the following items:
•
1 unopened cartridge
•
3.0 mL syringe and fill needle
•
Vial of Humalog
®
or NovoLog
®
insulin
•
Alcohol prep swab
•
1 new infusion set
•
Infusion set Instructions for Use
NOTE: Removing the Cartridge
Do NOT remove the used cartridge from the
pump during the load process until prompted
on the t:slim X2 Pump screen.
The illustration identifies the
Luer-lock and insulin fill port used in
the cartridge filling process.
Luer-lock
Tail of the
Cartridge
Insulin
Fill Port
Summary of Contents for t:slim X2
Page 2: ......
Page 14: ...Section 1 Before You Begin...
Page 15: ...Chapter 1 Introduction...
Page 19: ...Chapter 2 Important Safety Information...
Page 31: ...Chapter 3 Getting to Know Your t slim X2 Pump...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profiles...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim X2 Pump Settings...
Page 109: ...Chapter 12 t slim X2 Pump Info and History...
Page 111: ...Chapter 13 t slim X2 Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim X2 Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim X2 Pump Alerts...
Page 143: ...Chapter 16 t slim X2 Pump Alarms...
Page 157: ...Chapter 17 t slim X2 Pump Malfunction...
Page 160: ...Section 5 Living With and Caring for Your t slim X2 Pump...
Page 161: ...Chapter 18 Lifestyle Issues and Travel...
Page 165: ...Chapter 19 Taking Care of Your t slim X2 Pump...
Page 168: ...Section 6 Technical Specifications and Warranty...