Chapter 14 – User Settable Alerts and Alarms
14.1 Low Insulin Alert
Your t:slim X2 Pump keeps track
of how much insulin remains in the
cartridge and alerts you when it is low.
The default for this alert is pre-set to
20 units. You can set this alert setting
anywhere between 10 and 40 units.
When the insulin amount reaches the
set value, the Low Insulin Alert beeps/
vibrates and appears on the screen.
After the alert is cleared, the low insulin
indicator (a single red bar on the insulin
level display on the Home Screen
appears).
1.
From the Home Screen,
tap
OPTIONS
.
2.
Tap
My Pump
.
3.
Tap
Alert Settings
.
4.
Tap
Pump Alerts
.
5.
Tap
Low Insulin
.
14.2 Auto-Off Alarm
Your t:slim X2 Pump can stop insulin
delivery and alert you (or whoever is
with you) if there has been no interac-
tion with the pump within a specified
period of time. The default for this
alarm is pre-set to 12 hours. You can
set it anywhere between 5 and 24
hours, or off. This alarm notifies you
that there has been no interaction with
the pump in the specified number of
hours and the pump will shut down
after 30 seconds.
When the number of hours since you
have pressed the Screen On/Quick
Bolus Button and tapped any inter-
active screen option or delivered a
Quick Bolus passes the set value, the
Auto-Off Alarm beeps and appears on
the screen, and insulin delivery stops.
1.
From the Home Screen,
tap
OPTIONS
.
2.
Tap
My Pump
.
3.
Tap
Alert Settings
.
4.
Tap
Pump Alerts
.
6.
Using the onscreen keypad, enter
the number of units (from 10 to
40 units) that you want the Low
Insulin Alert value to be set, and
tap
DONE
.
7.
Tap
SAVE
when all changes
are complete.
To Respond to the Low Insulin Alert
To clear the alert, tap
CLOSE
.
LOW INSULIN ALERT
Cartridge level:
Less than 20 u remaining.
CLOSE
Summary of Contents for t:slim X2
Page 2: ......
Page 14: ...Section 1 Before You Begin...
Page 15: ...Chapter 1 Introduction...
Page 19: ...Chapter 2 Important Safety Information...
Page 31: ...Chapter 3 Getting to Know Your t slim X2 Pump...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profiles...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim X2 Pump Settings...
Page 109: ...Chapter 12 t slim X2 Pump Info and History...
Page 111: ...Chapter 13 t slim X2 Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim X2 Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim X2 Pump Alerts...
Page 143: ...Chapter 16 t slim X2 Pump Alarms...
Page 157: ...Chapter 17 t slim X2 Pump Malfunction...
Page 160: ...Section 5 Living With and Caring for Your t slim X2 Pump...
Page 161: ...Chapter 18 Lifestyle Issues and Travel...
Page 165: ...Chapter 19 Taking Care of Your t slim X2 Pump...
Page 168: ...Section 6 Technical Specifications and Warranty...