Specification Type
Specification Details
Basal Delivery Accuracy at all Flow
Rates (tested per IEC 60601-2-24)
±5%
The Pump is designed to vent automatically when there is a pressure difference between inside the
cartridge and the surrounding air. In certain conditions, such as a gradual elevation change of 1,000
feet, the pump may not vent immediately and delivery accuracy can vary up to 15% until 3 units have
been delivered or elevation changes by more than 1,000 feet.
Bolus Delivery Accuracy at all Volumes
(tested per IEC 60601-2-24)
±5%
Patient Protection from Air Infusion
The pump provides subcutaneous delivery into interstitial tissue and does not deliver intravenous injections.
Clear tubing aids in detecting air.
Maximum Infusion Pressure Generated
and Occlusion Alarm Threshold
30 PSI
Frequency of Basal Delivery
5 minutes for all Basal Rates
Retention Time of Electronic Memory
when Internal Pump Battery is
Fully Discharged (including Alarm
Settings and Alarm History)
Greater than 30 days
Infusion Set used for Testing
Unomedical Comfort
™
Infusion Set
table continued on next page …
Summary of Contents for t:slim X2
Page 2: ......
Page 14: ...Section 1 Before You Begin...
Page 15: ...Chapter 1 Introduction...
Page 19: ...Chapter 2 Important Safety Information...
Page 31: ...Chapter 3 Getting to Know Your t slim X2 Pump...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profiles...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim X2 Pump Settings...
Page 109: ...Chapter 12 t slim X2 Pump Info and History...
Page 111: ...Chapter 13 t slim X2 Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim X2 Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim X2 Pump Alerts...
Page 143: ...Chapter 16 t slim X2 Pump Alarms...
Page 157: ...Chapter 17 t slim X2 Pump Malfunction...
Page 160: ...Section 5 Living With and Caring for Your t slim X2 Pump...
Page 161: ...Chapter 18 Lifestyle Issues and Travel...
Page 165: ...Chapter 19 Taking Care of Your t slim X2 Pump...
Page 168: ...Section 6 Technical Specifications and Warranty...