Customer Technical Support:
(877) 801-6901
22
Chapter 2 – Important Safety Information
under delivery of insulin. This can cause
very low or very high blood glucose.
ALWAYS
twist the Luer-lock con-
nection between the cartridge tubing
and the infusion set tubing an extra
quarter of a turn to ensure a secure
connection. A loose connection can
cause insulin to leak, resulting in under
delivery of insulin. This can cause high
blood glucose.
DO NOT
disconnect the Luer-lock
connection between the cartridge
tubing and the infusion set tubing. If the
connection comes loose, disconnect
the infusion set from your body before
tightening. Failure to disconnect before
tightening can result in over delivery
of insulin. This can cause low blood
glucose.
DO NOT
remove or add insulin from
a filled cartridge after loading onto the
pump. This will result in an inaccurate
display of the insulin level on the Home
Screen and you could run out of insulin
before the pump detects an empty
cartridge. This can cause very high
blood glucose, or Diabetic Ketoacido-
sis (DKA).
DO NOT
deliver a bolus until you have
reviewed the calculated bolus amount
on the pump display. If you dose an in-
sulin amount that is too high or too low,
this could cause very high or very low
blood glucose. You can always adjust
the insulin units up or down before you
decide to deliver your bolus.
DO NOT
allow small children (either
pump users or non-users) to ingest
small parts, such as the rubber USB
port cover and cartridge components.
Small parts could pose a choking
hazard. If ingested or swallowed, these
small component pieces may cause
internal injury or infection.
The system includes parts (such as the
USB cable and infusion set tubing) that
could pose a strangulation or asphyxi-
ation hazard. Always use the appropri-
ate length of infusion set and arrange
cables and tubing to minimize the risk
of strangulation.
ENSURE
that these
parts are stored in a secure place when
not in use.
For patients who do not self-manage
their disease, the Feature Lock function
should ALWAYS be ON when the
pump is not being used by a caregiver.
The Feature Lock function is intended
to prevent inadvertent button presses
that may lead to insulin delivery or
changes in the pump settings. These
changes can potentially lead to hypo-
glycemic or hyperglycemic events.
For patients whose insulin adminis-
tration is managed by a caregiver,
DO
turn off the Quick Bolus feature to
avoid inadvertent bolus delivery. With
the Feature Lock turned on, the Quick
Bolus feature is disabled. Inadvertent
button presses or tampering with
the insulin pump could result in over
delivery or under delivery of insulin. This
can cause very low or very high blood
glucose which could result in serious
injury or death.
Summary of Contents for t:slim X2
Page 2: ......
Page 14: ...Section 1 Before You Begin...
Page 15: ...Chapter 1 Introduction...
Page 19: ...Chapter 2 Important Safety Information...
Page 31: ...Chapter 3 Getting to Know Your t slim X2 Pump...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profiles...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim X2 Pump Settings...
Page 109: ...Chapter 12 t slim X2 Pump Info and History...
Page 111: ...Chapter 13 t slim X2 Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim X2 Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim X2 Pump Alerts...
Page 143: ...Chapter 16 t slim X2 Pump Alarms...
Page 157: ...Chapter 17 t slim X2 Pump Malfunction...
Page 160: ...Section 5 Living With and Caring for Your t slim X2 Pump...
Page 161: ...Chapter 18 Lifestyle Issues and Travel...
Page 165: ...Chapter 19 Taking Care of Your t slim X2 Pump...
Page 168: ...Section 6 Technical Specifications and Warranty...