Key Pump Featur
es
2
Chapter 5 – Infusion Site Care and Loading Cartridge
Instructions for Drawing Insulin from Vial into Syringe
PRECAUTION
ALWAYS
remove all air bubbles before beginning insulin delivery. Ensure there are no air bubbles when
drawing insulin into the filling syringe, hold the pump with the white fill port pointed up when filling
the tubing, and ensure that there are no air bubbles in the tubing when filling. Air in the system takes
space where insulin should be and can affect insulin delivery.
The fill estimate displayed on the pump is the amount of insulin available for deliv-
ery. It does not include insulin needed to fill the tubing (up to 30 units) and a small
amount of insulin that is not available for delivery. When filling the syringe, add ap-
proximately 45 units to the amount of insulin you want available for delivery.
For example, the pump requires a minimum of 50 units available for delivery after
fill tubing has been completed. Fill the syringe with approximately 95 units to have
enough to fill your tubing and still have 50 units available for delivery.
1.
Inspect the needle and syringe package for any signs of damage. Discard any
damaged product.
2.
Wash your hands thoroughly.
3.
Wipe the rubber septum of the insulin vial with an alcohol swab.
4.
Remove the needle and syringe from their packaging. Securely twist needle
onto syringe. Safely remove protective cap from needle by pulling outward.
5.
Draw air into syringe up to the amount of insulin desired (see image A).
6.
With insulin vial upright, insert needle into vial. Inject air from syringe into vial.
Maintain pressure on syringe plunger (see image B).
Summary of Contents for t:slim X2
Page 2: ......
Page 14: ...Section 1 Before You Begin...
Page 15: ...Chapter 1 Introduction...
Page 19: ...Chapter 2 Important Safety Information...
Page 31: ...Chapter 3 Getting to Know Your t slim X2 Pump...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profiles...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim X2 Pump Settings...
Page 109: ...Chapter 12 t slim X2 Pump Info and History...
Page 111: ...Chapter 13 t slim X2 Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim X2 Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim X2 Pump Alerts...
Page 143: ...Chapter 16 t slim X2 Pump Alarms...
Page 157: ...Chapter 17 t slim X2 Pump Malfunction...
Page 160: ...Section 5 Living With and Caring for Your t slim X2 Pump...
Page 161: ...Chapter 18 Lifestyle Issues and Travel...
Page 165: ...Chapter 19 Taking Care of Your t slim X2 Pump...
Page 168: ...Section 6 Technical Specifications and Warranty...