Customer Technical Support:
(877) 801-6901
73
Key Pump Featur
es
2
Chapter 5 – Infusion Site Care and Loading Cartridge
5.6
Setting Site Reminder
This section describes how to set the
Site Reminder after you fill the cannula.
To set the Site Reminder without filling
the cannula, from the Home Screen,
tap OPTIONS, tap Load, tap Site
Reminder then follow the instructions
below.
PRECAUTION
DO NOT
change your infusion set before
bedtime or if you will not be able to test your
blood glucose 1–2 hours after the new infusion
set is placed. It is important to confirm that the
infusion set is inserted correctly and delivering
insulin. It is also important to respond quickly
to any problems with the insertion to ensure
continued insulin delivery.
1.
Tap SAVE if correct. Tap
Edit
Reminder
if settings need to
be changed.
2.
Tap
Remind Me In
and select the
number of days (1–3).
Z
Z
The default for the Site Reminder is
set for 3 days
3.
Tap
Remind Me At
. Use the
onscreen keypad to enter time
and tap
DONE
.
4.
Tap
Time of Day
to change AM
or PM. Tap
DONE
.
5.
Verify Site Reminder is set
correctly and tap
SAVE
.
Z
Z
“Setting Saved” screen
is displayed.
6.
Load screen is displayed.
Tap
DONE
.
Z
Z
A reminder to test BG in 1 to 2
hours will display.
7.
Tap
RESUME
.
NOTE: First Time Use
If this is the first time using your t:slim X2
Pump and a Personal Profile has not been
defined, a screen will notify you that a
profile must be activated to resume insulin.
Tap CLOSE.
8.
RESUMING INSULIN
screen is
temporarily displayed.
Summary of Contents for t:slim X2
Page 2: ......
Page 14: ...Section 1 Before You Begin...
Page 15: ...Chapter 1 Introduction...
Page 19: ...Chapter 2 Important Safety Information...
Page 31: ...Chapter 3 Getting to Know Your t slim X2 Pump...
Page 54: ...Section 2 Key Pump Features...
Page 55: ...Chapter 4 Getting Started...
Page 61: ...Chapter 5 Infusion Site Care and Loading Cartridge...
Page 75: ...Chapter 6 Personal Profiles...
Page 85: ...Chapter 7 Bolus...
Page 93: ...Chapter 8 Stop Resume Insulin...
Page 96: ...Section 3 Additional Pump Operations...
Page 97: ...Chapter 9 Temporary Basal Rate...
Page 101: ...Chapter 10 Quick Bolus...
Page 105: ...Chapter 11 t slim X2 Pump Settings...
Page 109: ...Chapter 12 t slim X2 Pump Info and History...
Page 111: ...Chapter 13 t slim X2 Pump Reminders...
Page 117: ...Chapter 14 User Settable Alerts and Alarms...
Page 120: ...Section 4 t slim X2 Pump Safety Alerts and Alarms...
Page 121: ...Chapter 15 t slim X2 Pump Alerts...
Page 143: ...Chapter 16 t slim X2 Pump Alarms...
Page 157: ...Chapter 17 t slim X2 Pump Malfunction...
Page 160: ...Section 5 Living With and Caring for Your t slim X2 Pump...
Page 161: ...Chapter 18 Lifestyle Issues and Travel...
Page 165: ...Chapter 19 Taking Care of Your t slim X2 Pump...
Page 168: ...Section 6 Technical Specifications and Warranty...