- 35 -
Error messages indicating bad test parameters or operational errors
(1) [NO
RECEIVER!!]
No graduated cylinder is set before the start of the test.
(2) [BATH
TEMP?]
The condenser block is not at the set temperature ± 6
°
C at the start of a test.
(3) [M.S FLWR POSITION]
A test has started when the level follower sensor is not at the upper fixed position.
(4) [HF
POINT?]
The final heating start point is not set correctly. The setting range is from 80% to 95%.
If started with a point outside this range, an error is generated.
(5) [THERMOFUSE]
The sample caught fire and the thermofuse was blown.
If the thermofuse is blown, the fire buzzer sounds and the shutter closes.
If the nitrogen pipe is connected to the fire containment nitrogen port ((8) in Fig.1:
Model AD-6 on Page 11), the solenoid valve opens automatically to discharge
nitrogen into the heater block to help controllig the fire.
If a test is excuted with shutter closed, stove section will be heated abnormally, and
then thermofuse will be blown. The fire containment shutter must be lifted/fixed to
the shutter hook during a test.
The above errors are attributable to bad settings or operation. Eliminate the causes of any
errors and repeat the test.