- 20 -
5 Starting a Test
5.1
Starting a Test
Press the START key.
(
4
)
Note (
4
)
The tester will not start if the starting conditions have not been satisfied, for example,
if the
condenser bath is not within ±6
°
C of the set temperature (see point (2) in
Section 10.4: Error Status and Faults on Page 34).
Do not start a test with shutter closed, thermofuse blows
(1) Adjust the photo levels of the initial boiling point sensor (IBP) and the fluid level follower
sensor (MS).
During photo level adjustment, the following message is displayed:
<PHOTO ADJUSTMENT>
IBP OUT 80
IN 3800
MS OUT
80 IN 3800
(2)
After photo level adjustment, the fluid level follower sensor begins to go down.
During the descent, the following message is displayed.
LEVEL FOLLOWER DESCENDING
(3) When the level follower sensor has detected the zero point of the graduated cylinder, the
RUNNING screen shown in Fig. 6 appears on the display and the RUN lamp lights. If the
graduated cylinder is not in the receiver compartment, a problem occurs and the test is
interrupted (see (1) in Section 10.4: Error Status and Faults on Page 34).
< RUNNING >
IBP
c
s
MODE(01) GASOLIN(300)
5
c
s
10
c
%
DIST VOL 0.0 (%)
20
c
%
DIST TEMP 25.0 (
o
C) 30 c
%
DIST TIME 0 sec
40
c
%
50
c
%
HEATER OUT 50 (V )
60
c
%
BATH TEMP 1.0 (
o
C) 70 c
%
REC TEMP 15.0 (
o
C) 80 c
%
HEAT TEMP 25 (
o
C) 90 c
%
95
c
RE CORR 0.1(%)
97
c
PREV.RESULT OFF
E.P
c
s
Temperatur
Dist time/rate
Fig. 6 RUNNING Screen (Testing)
If the FUNCTION key is pressed during testing, the Distillation Curve screen shown in Fig. 7
on Page 21 is displayed. Press the key once to view the distillation curve during testing and
once again to return to the RUNNING screen shown in Fig. 6.