- 31 -
(d) When entering several mode names, repeat steps (a) to (c).
9.3
Manual Heater Control
The following screen appears when a test is started with Manual heater control “ON”.
<MANUAL HEAT >
IBP 35.0c 350s
MODE(01) GASOLIN(300)
5 50.0c 70s
10
60.0c
4.5%
DIST VOL 97.5 (%)
20 75.0c 4.5%
DIST TEMP 25.0 (
o
C) 30
85.0c
4.5%
DIST TIME 300 sec
40 95.0c 4.5%
50
110.0c
4.5%
HEATER OUT
5
0 (V )
60 120.0c 4.5%
BATH TEMP 1.0 (
o
C) 70
130.0c
4.5%
REC TEMP 15.0 (
o
C) 80
140.0c
4.5%
HEAT TEMP 25 (
o
C) 90
150.0c
4.5%
95
170.0c
RE CORR 0.1(%)
97 185.0c
PREV.RESULT OFF
E.P 205.0c 250s
Fig.13 Manual Heater Control Run Screen
The test starts with 50% heater output. Operator can adjust heater output value by pressing
Up-Down keys.
Operator can change heater control method (Manual/Auto) by pressing START key during a
test. For example, operator wants to control heater output from 80 to 90% recovery, push
START key after starting a test, and then push START key again when recovery reaches to
80%, and then control heater output manually. Push START key again when recovery reaches
to 90%, the AD-6 starts to control heater output automatically until test end.
(Manual/Auto switching by START does not affect the setting of “Manu.Heat ON/OFF in the
Test Condition screen.)