- 18 -
4 Prior
to
Testing
!
Warning: Do not use flasks having a deformed bottom, fine
scratches or cracks, as these may break and cause a
fire during testing.
!
Warning: To prevent a sample from accidentally catching alight,
adhere to the following three precautions:
·
Do not leave the laboratory unattended during
testing.
·
Ensure that a fire extinguisher is located nearby.
·
Do not place any inflammable substances near the
tester.
!
Caution: The
standard
graduated
cylinder contains a Viton
O-ring. Do not use acetone, methyl ethyl ketone, acetic
ethyl, or any other solvent, as solvents may cause the
ring to inflate and damage the graduated cylinder.
※
Optional graduated cylinder with Perfluoroelastomer O-ring is available
!
Caution: Use boiling chips (Small stones or small porous
ceramics) to avoid irregular boiling.
4.1
Starting the Tester
Turn on the main power switch. The test unit comes on and displays the screen shown in Fig. 5.
< STANDBY 1 >
BAROMETER 1013 hPa
1 GASOLIN(300)
INIT TEMP1 250.0
o
C
2
KEROSIN(300)
INIT TEMP2 450.0
o
C
3
NAPHTHA(300) HEAT/FREEZ 95.0 %
4
KEROSIN(400)
FINAL H (AUTO) + 0.0 %
5
GAS-OIL(400)
TEMP RANGE 300
6
RESIDUE 1.0 %
7
DIST.RATE 4.5 %
8
SET BATH 0.5
o
C
9
SET REC 15.0
o
C
10
11
START TEMP 25.0
o
C
12
BATH TEMP 12.3
o
C
13
REC TEMP 18.0
o
C
(Mode Selection) (Mode Parameters)
Fig. 5 STANDBY Screen